Patient Bedside Monitor VNP9000V
◆ 12" color TFT display, 800x600 dpi;
◆ Separated Alarm and beep sound seting;
◆ Separated module: Main board+Parameter module(ECG, NIBP, SpO2);
◆ Display brightness adjustable;
◆ Real time S-T segment analysis, and ARR analysis;
◆ Sp02 Pitch Tone Variation and Drug Dose Calculation;
◆ Multi-display: Standard, Large font, Trend Coexis, OxyCRG dynamic
◆ Standby mode for Energy saving and suspend monitoring;
◆ 7-lead ECG waveforms display in phase;
◆ Capture dynamic waveforms and history waveforms review;
◆ VGA output;
◆ Built-in rechargeable lithium battery 4 hours, and battery volume checking.
◆ Efficient resistance to interference of defibrillator and HF knife;
◆ SD card memory and USB data output;
◆ TCP/lP networking platform enable abundant future upgrade& CMS
◆ Complete and flexible mounting solution for transportation and in-patient monitoring.
◆ Standard parameters: 5-lead ECG,SpO2, NIBP, RESP, TEMP, PR;
◆ Optional: Touch Screen, Printer, 3/12- lead EG, SD card, VGA output, 2/4-BP, EtC02, AG module, Masimo/Nellcor SpO2, Wall mount, Trolly.
◆ STANDARD PARAMETERS
◆ ECG:
Lead mode: 5-lead(R,L,F,N,C)
Lead selection: l, ll, lll, avR, avL,avF, V
Waveform: 3 and 7channel selectable
Gain selection: 0.5mm/mv, 1mm/mv, 2mm/mv
Sweep speed: 12.5mm/s; 25mm/s; 50mm/s
Heart rate range: Adult: 15~300bpm; Neonate:/pediatric:15~350bpm
Accuracy: +l bpm or +1%,whichecer is greater
Resolution: 1bpm
Filter: surgery mode: 1~20Hz
Monitor model: 0.5~40Hz
Diagnostic mode: 0.05~130Hz
Scaling signal: 1mv,+3%
Protection: withstand 4000VAC/50 voltage
isolation against electrosurgical
interference and defibrillation
Alarm range: 15~350bpm
◆ SPO2
Measurement range: 0~100%
Resolution: 1%
Accuracy: +2%(70-100%);0-69% unspecified
Alarm range: 0~100%
Pluse Rate: range:20~300bpm
Resolution: 1bpm
Error: +1bpm or +2%,whichever is greater
◆ NIBP
Method: Digital Automatic oscillometric
Operation mode : Manual/Automatic/continuous
Auto measurement time: Adjustable(1~480min)
Measurement Unit: mmHg/Kpa selectable
Measurement types: Systolic, Diastolic, Mean
Measurement range:
Range of Systolic pressure:
Adult: 40~270mmHg
Pediatric: 40~220mmHg
Neonate: 40~135mmHg
Range of mean pressure:
Adult: 20~235mmHg
Pediatric: 20~165mmHg
Neonate: 20~110mmHg
Range of diastolic pressure:
Adult: 10~215mmHg
Pediatric: 10~150mmHg
Neonate :10~100mmHg
Over-pressure protection: Double safety protection
Resolution: 1 mmHg
Alarm: Systolic.Diastolic,Mean
◆ RESP
Method: Thoracic impedance
Measurement range: Adult:7~120rpm;
Neonate/Pediatric:7~150rpm
Apnea alarm: YES,10~40s
Resolution: 1rpm
Accuracy: +2rpm
◆ TEMP
Compatible probe: YSI or CYF
Measurement range: 5~50℃
Resolution: 0.1℃
Accuracy: +0.1℃
Refreshing time: about 1
Average measuring time: <10s
◆ Optional Parameters
IBP (Optional)
IBP Waveform: 2-Channel
Max 8 channels
Measurement Range: -50-350mmHg
Resolution: 1mmHg
Unit: mmHg, kPa
Accuracy:
Static: ±1mmHg or 2%, whichever is greater
(exclusive of transducer)
Dynamc: ±4mmHg or 4%, whichever is greater
Sensitivity of transducer: 5 uV/V/mmHg,+2%
Impedance of transducer: 300~3000 0hm
Transducer sites: ART,PA,LAP,RAP,CVP,ICP,P1,P2
Selection of measurement range:
ART:0~+ 350mmHg
PA:-10~+ 120mmHg
CVP/RAP/LAP/ICP:-10~+40mmHg
P1/P2:-50~+ 360mmHg
EtC02(Sidertream CO2,0ptional)
Measurement range: 0~99mmHg
Accuracy: +2mmHg(0~40mmHg)
Sampling Range: 100ml/min
Sampling rate accuracy: 15%
Respiration rate: 0~120rmp
Respiration accuracy: +2rmp(0~70rmp)
+5rmp(>70rmp)
Respiration time: <240msec (10% to 90%)
Delay time: <2s
EtCO2 (Mainstreaam CO2,Optional)
Method: Infrared Spectrum
Range: 0.0-10%(0~76%)
Resolution: 1mmHg (0.1%)
Accuracy: <5%(+4.0 mmHg)
0r<10%(of Readings)
Suntech (Suntech):
Technique: Oscillometric
Patient Applications: Adult Pediatric, Neonate
Systolic Range: 40-265mmHh
Diatolic Range: 20-200mmHh
MAP Range: 27-222mmHh
Pulse Range: 25-300BPM
Transducer Accuracy: ±3mHg up to full range
Operation Conditions: 0℃Cto 50'℃,15%to 95% non-condensing RH
Storage Conditions: -20Cto 65℃,15%to 95% non-condensing RH
Recorder
Build-in, thermal array
Plethysmogram caveform: 2 channels
Record mode: manual, on alarm, time-defined
Recording width: 50mm
Printing speed: 50mm/s
Recording type: Frozen waveform record
NlBP recall record Trend table record Alarm record Fixed -time record
◆ BROCHURE
|
Patient Bedside Monitor VNP9000V |
File Type: | File Size: |
◆ FEATURES
Solutions & Products